Features

  • Two channel mass flow control electronics unit
  • Uses industry standard mass flow control valves
  • Simultaneous display of read & set on both valves
  • Analog and serial data outputs
  • Fully computer controllable

DESCRIPTION

The MFC-2 provides the precise, stable power and control signals needed to control two independent mass flow control valves or flow meters in a compact easy-to-use package. The MFC-2 is well suited for standalone or computer controlled applications where accurate flow control is vital, such as respirometry, process control or two-component gas mixing. Operation is intuitive and flexible, with all functions accessible via high quality rotary encoders.

Conventional flow meters and control valves are temperature- and pressure-dependent, making them unsuitable for critical work.  In contrast, temperature and pressure do not affect mass flow control valves, which deliver a closely-regulated output continuously corrected to STP.

The two line alphanumeric display simultaneously shows the set and read values for attached mass flow control valves or meters. Two analog outputs allow the flow rates of attached mass flow controllers and mass flow meters to be read by data acquisition systems.  A serial output allows logging of set-points and readings, and can also be used to control the unit.

The compact MFC-2 is highly accurate and will not contribute significant error to any mass flow control system.  It runs on noncritical 9 – 18 VDC, and includes a universal power supply.  Adjustments and settings are permanently retained in the absence of power.

specs

Controller/Meter Connectors: Two standard 15-pin D-sub female connectors; cable pin-out provided
Flow Meter Analog Output: BNC, independent output for each mass flow device; 0-5 VDC for 0-100% full scale flow
Flow Control Outputs: Independent 0 – 5 VDC for each device
Power Outputs: Internally generated, isolated supplies of +15.0 ± 0.25V and -15.0 ± 0.25V at up to 350 mA each per device; isolated from input power; can be used with mass flow devices that have +12 VDC, +15 VDVC or +/- 15 VDC input power requirements
Power Requirements: 12 – 24 VDC; universal adapter supplied
Panel Display: Digital, 2 x 16 alphanumeric LCD
Readout: Set and read information for both channels
Serial Output: RS232-8N1; baud rate 115200; both polled and broadcast modes available
Dimensions (W x D x H): 6 x 6 x 3 in. (15.2 x 15.2 x 7.6 cm)
Weight: 3 lb (1.4 kg)